Technical data

8510C Service Quick Reference Guide 2-3
Instrument Compatibility
Upgrading an 8360 Source
Upgrading an 8360 Source
All 8360 series instruments are compatible, but not all allow use of quick-step, test-port
power flatness correction, receiver calibration, and power domain functions without
upgrading them. Refer to Table 2-1 for upgrades required for these functions.
8510C revision C.07.00 or greater firmware requires use of an 8360 series source for
complete compatibility. Revision C.06.xx firmware does not support receiver calibration or
power domain functions.
Table 2-1 8360 Upgrade Summary
Agilent Model Serial Prefix Required for Test Port Flatness
Correction
4
or Receiver Cal
6
Required for
Quick Step
5
83630A
No modification required
1
83650A All
83651A
< 3103A
83601A upgrade kit
2,5
83621A 3103A
08360-60167 firmware kit
5
83631A 3104A to 3111A
08360-60201 firmware kit
5
3112A
No modification required
1
83620A
83622A 3103A
08360-60167 firmware kit
4
note 3
83623A 3104A to 3111A
08360-60201 firmware kit
4
note 3
83624A 3112A to 3144A
No modification required
1
note 3
83640A 3145A
No modification required
1
83642A
1. Fully compatible at time of shipment
2. Includes installation
3. Quick Step cannot be retrofitted to these models.
4. 8360 series requires firmware 23 Oct. 90.
5. 8360 series requires firmware 06 Mar. 91.
6. Receiver Cal requires 8510C rev. C.07.00 or greater.