User`s guide
9
Command Description
Setup source parameters
[PRESET] Reset source to a known state
POWER [POWER LEVEL] 5 dBm Set test port power to +5 dBm (
P
0 max
-
P
path loss
)
Create frequency list
FREQUENCY[MENU] Access the frequency softkey menu
[more] Display next softkey menu
[Fltness Menu] Access the frequency list softkey menu
Enter List [Freq] 5 GHz Enter 5 GHz as the first frequency
Entering a frequency automatically sets the offset and dwell to 0 dB and 10 ms, respectively
Enter 18, 13, 11, and 20 GHz from the keypad
Access user flatness correction menu
POWER [MENU] Access power softkey menu
[more] Display next softkey menu
[Fltness Menu] Access flatness softkey menu
Clear flatness array
[Delete Menu] Access delete array softkey menu
Delete [All] Clear array
MENU SELECT [PRIOR] Leave delete menu
Display previous softkey menu
[more] Display next softkey menu
[more] Display next softkey menu
[Copy List] Copy the frequency list into the correction table in a sequential order
[Freq Follow] Set source to the active correction frequency in the flatness array
(i.e when the arrow points to the 5 GHz correction data point in the
correction table, the source is also set to a CW frequency of 5 GHz)
[more] Display next softkey menu
Calibrate the power meter to the power sensor in use
Connect power meter sensor to test port
Enter the cal factor for first correction frequency (5 GHz)
Enter correction data into table
POWER [FLTNESS ON/OFF] Enable User Flatness Correction; power meter will display
corrected data
Enter [Corr] Enter correction data - rotate knob until power meter displays
desired power level (5 dBm)
Allow entry of next frequency correction
Enter in the power meter the sensor's cal factor for the next frequency in the array
Repeat until each correction has been entered
[SAVE]n Save the source parameters including the correction table in an
internal register
Remove power meter sensor from test port and reconnect device
Activate List Mode
SWEEP [MENU] Access sweep softkey menu
Sweep Mode [List] Select frequency list mode
➮