User`s guide
2
The 8360 series synthesized sweepers provide
extremely flat power at your test port, for testing
power sensitive devices such as amplifiers, mixers,
diodes or detectors. The user flatness correction
feature of the 8360 synthesized sweepers compen-
sates for attenuation and power variations created by
components between the source and the test device.
User flatness correction allows the digital correction
of up to 801 frequency points (1601 points via GPIB),
in any frequency or sweep mode (i.e., start/stop, CW,
power sweep, etc.). Using a power meter to calibrate
the measurement system, as shown in figure 1, a
table of power level corrections is created for the
frequencies where power level variations or loss occur
(see figure 2). These frequencies may be sequential
linear steps or arbitrarily spaced. To allow the
correction of multiple test setups or frequency
ranges, the user may save as many as eight different
measurement setups (including correction tables)
in the internal storage registers of the 8360.
This note illustrates how to utilize the user flat-
ness correction feature by providing step-by-step
instructions for several common measurement
examples.
Introduction
Figure 1. Basic system configuration using an 437B power meter
for automatic user flatness correction data entry
Figure 2. 8360 displaying the user flatness correction table