User`s guide

14
Example 5:
Computer controlled measurements
with user flatness corrections
This section focuses on computer controlled meas-
urement systems. The following program interrogates
the source and a 437B power meter for frequency
and power information respectively. The source
(83620B) is programmed to sweep from 2 to 20 GHz,
with user flatness corrections every 100 MHz and
+5 dBm leveled output power. It is assumed that
the path losses do not exceed 5 dBm and that the
437B power meter already has its power sensor’s
calibration factors stored in sensor data table 0. If
another power meter is utilized, the power sensor’s
calibration factors will have to be stored in a look-up
table. Modify the program to suit your particular
measurement requirements. Up to 801 points may
be entered in the user flatness correction table
with this program.
TMSL commands (see figure 11a) are used to set
up the source parameters and enter correction
frequencies and data into the user flatness correction
table. The TMSL commands are written in capitalized
and lower case letters. The capitalized portion of
each command illustrates the short form. Commands
may be written in either the short or the long form,
or any combination of the two. Refer to the 8360
User’s Guide for more details on TMSL commands.
Figure 11a. TMSL commands