User`s guide
Getting Started with Return Loss A Background to Return Loss Measurement
36 Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
A Background to Return Loss
Measurement
Measuring the Reflected Power from a
Component with Known Reflection Factor
First, attach a component with a known reflection factor in place of
the DUT, and measure the power reflected. This is component is called
the reflection reference.
Figure 29 Measuring the Power from a Component with a Known Reflection Factor
This measured power from the reflection reference is called P
Ref
.
The following two components are recommended for use as a
reflection reference:
• the Agilent 81610CC Reference Cable, or
• the Agilent 81000BR Back Reflector.
NOTE Note you can only measure the Front Panel Delta if you use a
Reflection Reference Cable.
The reflection factor for the component is called R
R
. Normally the
return loss for the component (RL
R
) is specified, but these values are
related:
Reflection
Reference
R
ref
M
ref
P
ref
Agilent 81610A
Reflectivity
P
in
P
r
5/ 10 5log– 10
3
3
-------log–==