1 Installing the Agilent 82357B
Connecting the 82357B to your PC
This section shows steps to connect the 82357B to a USB
port on your PC or to your PC via a USB Hub.
1 Connect to a USB Port. Make sure the PC is ON and
plug the 82357B USB cable into any available USB port
on your PC. Do not connect the 82357B to GPIB
instruments at this time.
Figure 1-4 Connecting the 82357B to Your PC
2 Observe the LEDs. Observe the LEDs on the 82357B for
at least 10 seconds. See Chapter 2, “Using the Agilent
82357B” for a description of the normal LED sequence
during an initial installation of the 82357B.
a Initially, only the red FAIL LED should be ON. After a
few seconds, all three LEDs should be ON. When all
three LEDs are ON, this shows that the 82357B has
been successfully installed.
b If all three LEDs are not ON after 10 seconds and all
Windows Plug- and- Play Manager activity has ceased,
STOP. See Chapter 3, “Troubleshooting the Agilent
82357B” for diagnostics information.