User`s guide

Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
68
10 Read the DVM and change the high voltage value on the display to match the
measured voltage.
For example, if the DVM reads 19.495 V, adjust the voltage to 19.495 V using the
knob and arrow keys.
11 Pressing (Calibrate) key saves the new voltage calibration constants, and
goes to the OVP calibration mode.
A ‘‘CALIBRATING’’ message appears for one second to indicate that the voltage
calibration is progressing and new voltage calibration constants of ‘‘SETUP 1’’ are
stored. Then, the display shows above message to indicate that the power supply is
ready for the OVP calibration.
If the calibration fails, a ‘‘DAC CAL FAIL’ or ‘‘ADC CAL FAIL’ message appears
for one second and the display shows the ‘‘CAL SETUP 1’’ for voltage calibration
again.
Current and OCP Calibration
12 Pressing (Calibrate) key saves the new calibration constants for OVP
circuit and goes to the current calibration mode.
A ‘‘CALIBRATING’’ message appears for about several seconds to indicate that the
OVP calibration is progressing and new calibration constants of ‘‘SETUP 2’’ are
stored. Then, the display shows the above message to indicate that the power supply
is ready for the current calibration.
V HI 19.495 V
CAL SETUP 2
CAL SETUP 3
Connect an appropriate shunt 0.01 Ω across the output terminals, and connect a
digital voltmeter across the shunt resistor for the current calibration.
Error
Calibrate
Error
Calibrate
Error
Calibrate
Error
Calibrate