User`s guide

Chapter 3 Calibration Procedures
Calibration Security Code
61
3
Calibration Security Code
This feature allows you to enter a security code (electronic key) to prevent accidental
or unauthorized calibrations of the power supply. When you first receive your power
supply, it is secured. Before you can calibrate the power supply, you must unsecure
it by entering the correct security code. A procedure to unsecure the power supply is
given on the following page.
The security code is set to ‘‘HP003633’’* or ‘‘HP003634’’** when the power
supply is shipped from the factory. The security code is stored in non-volatile
memory, and does not change when power has been off or after a remote interface
reset.
To secure the power supply from the remote interface, the security code may
contain up to 12 alphanumeric characters as shown below. The first character must
be a letter, but the remaining characters can be letters or numbers. You do not have
to use all 12 characters but the first character must always be a letter.
A _ _ _ _ _ _ _ _ _ _ _ (12 characters)
To secure the power supply from the remote interface so that it can be unsecured
from the front panel, use the eight-character format shown below. The first two
characters must be ‘‘H P’’ and the remaining characters must be numbers. Only
the last six characters are recognized from the front panel, but all eight characters
are required. To unsecure the power supply from the front panel, omit the ‘‘HP’’
and enter the remaining numbers as shown on the following pages.
HP _ _ _ _ _ _ (8 characters)
If you forget your security code, you can disable the security feature by adding a
jumper inside the power supply, and then entering a new code.
See the procedure on page 63.
*For Agilent E3633A Model **For Agilent E3634A Model