User`s guide
Chapter 3 Calibration Procedures
Constant Voltage (CV) Verifications
55
3
Load Transient Response Time
This test measures the time for the output voltage to recover to within 15 mV of
nominal output voltage following a load change from full load to half load, or half
load to full load.
1 Turn off the power supply and connect the output to be tested as shown in Figure 3-
1 with an oscilloscope. Operate the electronic load in constant current mode.
2 Turn on the power supply. Select 20V/10A* or 50V/4A** range, enable the output,
and set the display to the limit mode. When the display is in the limit mode, program
the current to the full rated value (10.0 A)* or (4.0 A)** and the voltage to the full
rated value (20.0 V)* or (50.0 V)**.
3 Set the electronic load to transient operation mode between one half of the output’s
full rated value and the output’s full rated value at a 1 kHz rate with 50% duty cycle.
4 Set the oscilloscope for ac coupling, internal sync, and lock on either the positive or
negative load transient.
5 Adjust the oscilloscope to display transients as shown in Figure 3-3. Note that the
pulse width (t2 - t1) of the transients at 15 mV from the base line is no more than 50
μsec for the output.
Figure 3-3. Transient Response Time
*For Agilent E3633A Model **For Agilent E3634A Model