User`s guide
Chapter 3 Calibration Procedures
Constant Voltage (CV) Verifications
54
CV PARD (Ripple and Noise)
Periodic and random deviations (PARD) in the output (ripple and noise) combine to
produce a residual ac voltage superimposed on the dc output voltage. CV PARD is
specified as the rms or peak-to-peak output voltage in the frequency range from 20
Hz to 20 Mhz.
• VRMS measurement techniques:
When measuring Vrms ripple and noise, the monitoring device should be plugged into
the front of the terminals at (B) in Figure 3-2. Use the coaxial cable and BNC female
to banana plug adapter to connect the monitor device to the power supply. To reduce
the measurement error caused by common mode noise, it is recommended that you
use a common mode choke between the cable and the BNC adapter. The common
mode choke is constructed by inserting coaxial cable into Ferrite Toroidal core. The
load resistor should be connected to the terminal at (A) in Figure 3-2. Twisted leads
between the load resistor and the power supply helps reduce noise pickup for these
measurements.
1 Turn off the power supply and connect the output to be tested as shown in Figure 3-
1 to an oscilloscope (ac coupled) between (+) and (-) terminals. Set the oscilloscope
to AC mode and bandwidth limit to 20 MHz. Connect a resistive load (2.0
Ω)* or
(12.5
Ω)** to the terminal at (A) in Figure 3-2.
2 Turn on the power supply. Select the 20V/10A* or 50V/4A** range, enable the output,
and set the display to the limit mode. When the display is in the limit mode, program
the current to the full rated value (10.0 A)* or (4.0 A)** and the voltage to the full
rated value (20.0 V)* or (50.0 V)**.
3 Check that the front panel CV annunciator remains lit. If not lit, adjust the load down
slightly.
4 Note that the waveform on the oscilloscope does not exceed the peak-to-peak limit
of 3 mV.
5 Disconnect the oscilloscope and connect the ac rms voltmeter in its place according
to the VRMS measurement techniques above. The rms voltage reading does not
exceed the rms limit of 0.35 mV* or 0.5 mV**.
*For Agilent E3633A Model **For Agilent E3634A Model