User`s guide
Chapter 3 Calibration Procedures
Measurement Techniques
49
3
Measurement Techniques
Setup for Most Tests
Most tests are performed at the front terminals as shown in the following figure.
Measure the dc voltage directly at the (+) and (-) terminals on the front panel.
Figure 3-1. Performance Verification Test Setup
Electronic Load
Many of the test procedures require the use of a variable load resistor capable of
dissipating the required power. Using a variable load resistor requires that switches
should be used to connect, disconnect, and short the load resistor. An electronic load,
if available, can be used in place of a variable load resistor and switches. The electronic
load is considerably easier to use than load resistors. It eliminates the need for
connecting resistors or rheostats in parallel to handle power, it is much more stable
than carbon-pile load, and it makes easy work of switching between load conditions
as is required for the load regulation and load response tests. Substitution of the
electronic load requires minor changes to the test procedures in this chapter.