User`s guide

Chapter 3 Calibration Procedures
Test Considerations
46
You can also enter calibration constants from the remote interface. Remote operation
is similar to the local front-panel procedure. You can use a computer to perform the
adjustment by first selecting the required setup. The calibration value is sent to the
power supply and then the calibration is initiated over the remote interface. The power
supply must be unsecured prior to initiating the calibration procedure. An example
program of Excel 97 for calibration over the GPIB interface is listed at the end of this
chapter.
For further details on programming the power supply, see chapters 3 and 4 in the
Agilent E3633A and E3634A Users Guide.
Test Considerations
To ensure proper instrument operation, verify that you have selected the correct
power-line voltage prior to attempting any test procedure in this chapter. See page 24
in chapter 2 for more information.
Ensure that all connections of terminals (both front panel and rear panel) are removed
while the power supply internal self-test is being performed.
For optimum performance verification, all test procedures should comply with the
following recommendations:
Assure that the calibration ambient temperature is stable and between 20°C and
30°C.
Assure ambient relative humidity is less than 80%.
Allow a 1-hour warm-up period before verification or calibration.
Use short cables to connect test set-ups.
Caution The tests should be performed by qualified personnel. During performance
verification tests, hazardous voltages may be present at the outputs of the power
supply.