User`s guide
Chapter 2 Quick Start
To Check the Rated Currents of the Power Supply
26
To Check the Rated Currents of the Power Supply
The following procedures check to ensure that the power supply produces its rated
current outputs with a short and properly responds to operation from the front panel.
For each step, use the keys shown on the left margins.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled (the
OFF annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the 8V* or
25V** annunciator turns on); and the knob is selected for voltage control.
2 Connect a short across (+) and (-) output terminals with an insulated test lead.
3 Enable the outputs.
The OFF annunciator turns off and the 8V* or 25V**, OVP, and OCP annunciators
turn on. The
CV or CC annunciator is lit depending on the resistance of the test lead.
The flashing digit can be adjusted by turning the knob. Notice that the display is in
the meter mode. “Meter mode’’ means that the display shows the actual output voltage
and current.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the Limit annunciator will be flashing). Adjust the
voltage limit to 1.0 volt to assure CC operation. The
CC annunciator will light.
5 Check that the front-panel ammeter properly responds to knob control.
Set the knob to the current control, and turn the knob clockwise or counter clockwise
when the display is in the meter mode (the
Limit annunciator is off). Check that the
ammeter responds to knob control and the voltmeter indicates nearly zero (actually,
the voltmeter will show the voltage drop caused by the test lead).
Power
Output
On/Off
Display
Limit
Voltage
Current
*For Agilent E3633A Model **For Agilent E3634A Model