User`s guide
Performance Tests Performance Tests
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
49
Test Setup
Perform the test until the power reading of the DUT shows:
for 81623B: -47.x dBm
for 81624B: -57.x dBm
for 81626B: -37.x dBm
for 81628B: -57.x dBm (without sphere)
1 Make sure that cable connector, detectors and adapters are clean.
2 Make sure that you perform this test in a temperature-controlled
environment with temperature fluctuations less than ±1°C.
3 Setup the equipment as shown in Figure 9 . Disable both attenuators
and enable the laser source, where the source wavelength is chosen to
1550 nm. If you are using the 81657A, move to the wavelength
parameter, [λ], press Enter, select the longer wavelength source
(1550 nm nominally), and press Enter.
4 Set the wavelength of both attenuators to the same wavelength as the
laser source.
5 Perform the following sub-procedure for both Power Meters:
a Move to the Power Meter channel.
b Move to the wavelength parameter, [λ], press Enter, enter the
wavelength of the laser source, and press Enter.
c Move to the calibration parameter, [CAL], press Enter, set the
calibration parameter to zero, and press Enter.
d Move to the averaging time parameter, [AvgTime], press Enter,
move to <100 ms>, and press Enter.
e Move to the power parameter, [P], press [Pwr unit], move to <dBm>,
and press Enter.
• Do not turn the laser off during the measurement!
• Clean all connectors carefully before you start with the measurement!
The linearity test checks the in-range linearity as well as the range
discontinuity. Therefore, two values are measured at the same power but
within different power ranges (range discontinuity) and two different
power values , one on the lower and the other on the upper limit of the
power range, are measured within the same power range (in-range
linearity).
NOTE
NOTE