User`s guide

Te st R e c ord Performance Tests
Agilent 81689A, 81689B and 81649A Tunable Laser Modules Users Guide, Third Edition 85
Agilent 81689A Performance Test
Page 5 of 8
Model Agilent 81689A Tunable Laser Report No. ________ Date_______
Measurement Uncertainty: ± 0.1 pm
Maximum Power Test
Measurement Uncertainty: ± 0.10 dB
Repeatability of 1620.000 nm
(= reference)
Measurement
Result
Initial Setting REF = nm
from 1570000 nm to REF nm
from 1580.000 nm to REF nm
from 1590.000 nm to REF nm
from 1600.000 nm to REF nm
from 1610.000 nm to REF nm
largest measured wavelength nm
smallest measured wavelength nm
Wavelength Repeatability nm
= largest measured wavelength - smallest measured wavelength
Specification 0.10 nm
Wavelength Setting Power Measured Minimum Specification
1570.000 nm dBm + 6.00 dBm
1575.000 nm dBm + 6.00 dBm
1580.000 nm dBm + 6.00 dBm
1585.000 nm dBm + 6.00 dBm
1590.000 nm dBm + 6.00 dBm
1595.000 nm dBm + 6.00 dBm
1600.000 nm dBm + 6.00 dBm
1605.000 nm dBm + 6.00 dBm
1610.000 nm dBm + 6.00 dBm
1615.000 nm dBm + 6.00 dBm
1620.000 nm dBm + 6.00 dBm