Power Stability
Follow the steps below to measure the power stability:
1 Set up the equipment as shown in Figure 12.
2 Move to the Tunable Laser channel of the Agilent 8164A/B
Lightwave Measurement System and press [Menu].
3 Set the menu parameters to the values shown in Table 5.
4 Set the wavelength and power for each Tunable Laser module to the
values given in Table 13.
5 Make sure the optical output is switched off.
6 Zero the power meter. From [Menu], select [Zero].
7 Press the key beside the laser output to switch on the laser.
8 Select the logging application. Press [Appl], select [Logging]
9 Within the logging application, set the power meter:
•Set λ, the wavelength, to the same as your Tunable Laser module, as
given in Table 13.
•Set Range to 0dB
Set Ref mode to Value
Set Samples to 4000
Set the Average Time to 200 ms
Set Range mode to common
Set Power Unit to dB
Set Ref to the value given in Table 13.
10 Start the Logging Application by pressing [Measure]
A measurement progress indicator is displayed.
11 When the measurement is completed, select [Analysis]
12 From the Statistics window, note:
the “max” value in the Maximum Deviation field of the test record,
the “min” value in the Minimum Deviation field of the test record,
Table 13 Wavelength and Power Settings for Power Stability Tests
Module Wavelength] Power [P]
Agilent 81689A 1540.000 nm 3.00 dBm
Agilent 81689B 1540.000 nm 0.00 dBm
Agilent 81649A 1570.000 nm 3.00 dBm