Specifications
82
Chapter 3: Testing Performance
Test Pod 1 in the 1250 Mb/s mode
necessary. See page 56.
20 Re-set the pulse generator frequency to 625 MHz plus the test margin (638
MHz).
21 In the Eye Finder window, align the blue bars vertically. See page 55.
22 Grab the blue bar for “Label1 (4 Channels)” and move it to the
recommended starting position you noted in the prior step.
23 Run Eye Finder again. Some eyes may close, but the eyes in the sampling
position you chose on page 81 should remain open. Realign stray blue bars
if necessary. See page 56.