Specifications

80
Chapter 3: Testing Performance
Test Pod 1 in the 1250 Mb/s mode
Test Pod 1 in the 1250 Mb/s mode
1 Disconnect the Pod 2 E5382A Flying Lead Probe Set’s bits 2, 6, 10, and 14
from the SMA/Flying Lead test connectors’ pin strip connectors.
The original E5382A Flying Lead Probe Set should still be connected to
Pod 1. The J clock on Pod 1 will be used for all tests.
2 Connect the Pod 1 E5382A Flying Lead Probe Set’s bits 6 and 14 to the
SMA/Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
3 Connect the Pod 1 E5382A Flying Lead Probe Set’s bits 2 and 10 to the
SMA/Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
4 On the 8133A pulse generator, in the PULSE setup for CHANNEL 2, press
the COMP button to return the outputs to normal.
5 Note that the signal on the oscilloscope has moved. Change the
oscilloscope’s horizontal position to 450 ps (or as required) to center the
measured pulse on the oscilloscope display.
NOTE: You can temporarily reduce the frequency on the pulse generator and
increase the horizontal scale on the oscilloscope to verify that you are
measuring the correct pulse. Then return the settings to normal.
6 Verify the DC offset and adjust it if necessary. See page 46.
7 Set the frequency of the pulse generator. The logic analyzer will be tested
using a double-edge clock. The test frequency is half the test clock rate
because data is acquired on both the rising edge and the falling edge of the
clock. Set the frequency to 625 MHz plus the frequency uncertainty of the
pulse generator, plus a test margin of 1%.
For example, if you are using an 8133A pulse generator, the frequency
accuracy is ±1% of setting. Use a test margin of 1%. Set the frequency to
625 MHz plus 2% (638 MHz).
8 Verify the oscilloscope Deskew and adjust if necessary. See page 47.
9 Adjust the measured pulse width from the pulse generator to 750 ps
(minus the test margin) as described on page 72.
10 In the logic analyzer “Setup and Trigger...” window, select the Sampling