Specifications

72
Chapter 3: Testing Performance
Test Pod 1 in the 800 Mb/s mode
Test Pod 1 in the 800 Mb/s mode
1 Disconnect the Pod 2 E5382A Flying Lead Probe Set’s bits 2, 6, 10, and 14
from the SMA/Flying Lead test connectors’ pin strip connectors.
The original E5382A Flying Lead Probe Set should still be connected to
Pod 1. The J clock on Pod 1 will be used for all tests.
2 Connect the Pod 1 E5382A Flying Lead Probe Set’s bits 6 and 14 to the
SMA/Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
3 Connect the Pod 1 E5382A Flying Lead Probe Set’s bits 2 and 10 to the
SMA/Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
4 On the 8133A pulse generator, in the PULSE setup for CHANNEL 2, press
the COMP button to return the outputs to normal.
5 Note that the signal on the oscilloscope has moved. Change the
oscilloscope’s horizontal position to 800 ps (or as required) to center the
measured pulse on the oscilloscope display.
6 Verify the DC offset and adjust it if necessary. See page 46.
7 Set the frequency of the pulse generator. The logic analyzer will be tested
using a double-edge clock. The test frequency is half the test clock rate
because data is acquired on both the rising edge and the falling edge of the
clock. Set the frequency to 400 MHz plus the frequency uncertainty of the
pulse generator, plus a test margin of 1%.
For example, if you are using an 8133A pulse generator, the frequency
accuracy is ±1% of setting. Use a test margin of 1%. Set the frequency to
400 MHz plus 2% (408 MHz).
8 Verify the oscilloscope Deskew and adjust if necessary. See page 47.
Adjust the measured pulse width to 750 ps
In this procedure, you will use the oscilloscope’s measurement markers to
measure the actual pulse width in the test setup. Then you will adjust the pulse
generator so that the measured pulse width is as specified.
9 Observe the 54845A oscilloscope display. Change the Channel 2 pulse