Specifications

64
Chapter 3: Testing Performance
Test Pod 2 in 200 Mb/s mode
Test Pod 2 in 200 Mb/s mode
1 Leave the first E5382A Flying Lead Probe Set connected to Pod 1 of the
logic analyzer. Remove the Pod 1 flying leads 2, 6, 10, and 14 from the
SMA/Flying Lead test connectors. Do not remove the flying leads that are
connected to CLK and CLK.
2 Connect the second E5382A Flying Lead Probe Set to Pod 2 of the logic
analyzer.
3 Connect the Pod 2 E5382A Flying Lead Probe Set’s bits 6 and 14 to the
SMA/Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
4 Connect the Pod 2 E5382A Flying Lead Probe Set’s bits 2 and 10 to the
SMA/Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
5 On the 8133A pulse generator, in the PULSE setup for CHANNEL 2, press
the COMP button to return the outputs to normal.
6 Note that the signal on the oscilloscope has moved. Change the
oscilloscope’s horizontal position to 800 ps (or as required) to center the
measured pulse on the oscilloscope display.
7 Verify the DC offset and adjust it if necessary. See page 46.
8 Deskew the oscilloscope if necessary. See page 47.
9 Readjust the pulse width from the pulse generator as measured on the
oscilloscope to 1.5 ns. See page 48.
10 In the logic analyzer’s Setup and Trigger window, Format tab, unassign all
pod 1 bits. A Warning window will appear stating that the trigger function
has become invalid. Select OK to close the Warning window.
11 Assign bits 2, 6, 10, and 14 of Pod 2.
12 Set the Pod 2 threshold to 1 volt (just as you did for Pod 1 on page 53).
13 In the “Setup and Trigger...” window, Format tab, use the scroll bar at the
bottom of the window to scroll to the left. Ensure that the J clock
threshold is set to Differential.
14 Re-establish the trigger function:
a In the logic analyzer’s Setup and Trigger window, select the Trigger tab,