Specifications

61
Chapter 3: Testing Performance
Test Pod 1 in 200 Mb/s mode
for marker...”.
1 Let the logic analyzer run repetitive for about 1 minute. If no error
message is displayed the test passes.
If the red error window “Error- Listing <1> Pattern NOT found...” appears
during the minute or so that it runs repetitively, then the logic analyzer
fails the test. Record the failure in the appropriate place on the test
record.
NOTE: Be sure that the black ground clip is making good contact with the ground
pin on the test connector.
2 After approximately 1 minute select the red stop button in the Listing
window.