Specifications

55
Chapter 3: Testing Performance
Configure the logic analysis system
d Enter “A” in the “Label 1 =” field.
Adjust sampling positions using Eye Finder
1 In the Setup and Trigger window, select the Sampling tab, then select the
“Sampling Positions...” button (which is in the Clock Setup area of the
window). The Eye Finder window will appear. Ensure that the “Sampling
Positions” tab is selected.
2 From the menu bar in the Eye Finder window, select Results.
3 Select “Remove All Eye Finder Data” if this selection is available (if it is
not ghosted). Select “Yes” to confirm that you want to remove the data.
4 Select Label1, then select Expand (if not already expanded).
5 If the blue bars in the Eye Finder display are not vertically aligned:
a Grab the right-most blue bar in the “Label1 (4 channels)” row with the
mouse pointer and move it all the way to the left and release the mouse
button. This will vertically align all of the blue bars.
6 Select the Sampling Positions tab in the Eye Finder window.
7 Using the mouse pointer, grab the blue bar for “Label 1 (4 channels)” and