Chapter 3: Testing Performance
Configure the logic analysis system
Configure the logic analysis system
1 Configure the Sampling settings.
a In the Agilent 16700-series logic analysis system’s System window,
select the 16760A logic analyzer icon, then select “Setup and Trigger”.
A Setup and Trigger window will appear.
b In the logic analyzer Setup and Trigger window, select the Sampling tab
(if it is not already selected).
c Under the Sampling tab, select State Mode.
d Under State Mode Controls, select the “200 Mb/s / 32 M State” mode.
e Under Clock Setup, at the bottom of the Clock J column, select the
clock edge field and set the J clock to “Both Edges”.
2 Configure the Acquisition settings.
a In the logic analyzer Setup and Trigger window, select the Trigger tab.
b Under the Trigger tab, select the Settings subtab.
c Select the Acquisition Depth field, then select 128K.
d Select the Trigger Position field, then select Start.