Specifications
46
Chapter 3: Testing Performance
Set up the test equipment
Set up the test equipment
Next, verify and adjust the pulse generator’s DC offset, deskew the oscilloscope,
and measure and adjust the pulse width.
Verify and adjust 8133A pulse generator DC offset
1 On the 54845A oscilloscope, select Measure from the menu bar at the top
of the display.
2 Select Markers...
3 In the Markers Setup window set marker “Ay” to 0.875 V, and set marker
“By” to 1.125 V.
4 Observe the waveforms on the oscilloscope display. If they are not
centered within the “Ay” and “By” markers, adjust the 8133A pulse
generator’s Channel 2 OFFSET until the waveforms are centered as well as
possible. (The resolution of the 8133A OFFSET setting is 20 mV.)