Specifications
45
Chapter 3: Testing Performance
Connect the test equipment for the minimum clock interval/minimum eye width test
NOTE: Be sure to use the black ground clip (supplied with the E5382A Flying
Lead Probe Set) and orient the leads so that the black clip is connected to
one of the SMA/Flying Lead connector’s ground pins!
6 Connect the E5382A Flying Lead Probe Set’s CLK lead to the SMA/Flying
Lead connector at the 8133A pulse generator’s Channel 1 OUTPUT
. Again,
be sure to use the black ground clip and orient the leads so that the black
clip is connected to ground.
7 Connect the E5382A Flying Lead Probe Set’s bits 6 and 14 to the SMA/
Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
8 Connect the E5382A Flying Lead Probe Set’s bits 2 and 10 to the SMA/
Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
Connect the 8133A Pulse Generator Output to the 54845A
Oscilloscope
9 Attach Male BNC to Female SMA adapters to Channels 1 and 2 on the
54845A oscilloscope.
10 Attach one end of an SMA cable to the Male BNC to Female SMA adapter
on Channel 1 of the oscilloscope.
11 Attach the other end of the SMA cable to the SMA/Flying Lead connector
at the Channel 2 OUTPUT of the 8133A pulse generator.
12 Attach one end of the other SMA cable to the Male BNC to Female SMA
adapter on Channel 2 of the oscilloscope.
13 Attach the other end of the SMA cable to the SMA/Flying Lead connector
at the Channel 2 OUTPUT of the 8133A pulse generator.