Specifications
41
Chapter 3: Testing Performance
Prepare the logic analysis system for testing
Some system CPU board tests may return a status of Untested because they require
user action. Procedures to do these tests are found in the Agilent Technologies
16700B-series Logic Analysis System Service Guide. For the purposes of testing the
Agilent 16760A module performance, running untested system CPU board tests are
not required. If these tests are not done, the Agilent 16760A performance test is not
affected.
d When the self-tests are complete, select Quit to exit the test menu.
e Record a PASS in the Performance Test Record if all module self-tests
pass.
4 Set up the logic analysis system.
a In the Session Manager window, select "Start Session" to re-launch a
logic analysis session.
b In the Logic Analysis System window, select the module icon, then
select Setup and Trigger. A Setup and Trigger window will appear.
c In the Setup and Trigger window, click on the Calibration tab. Follow
the instructions under the Calibration tab to perform an operational
accuracy calibration on the Agilent 16760A module.
Repeat the above step for each single-card Agilent 16760A module installed in the
mainframe that is being tested. If any calibration status returns "failed," then the
Agilent 16760A module requires repair. If the calibration status returns "passed" for
both pods, then record a PASS in the performance test record at the end of this
chapter.
d Proceed to the next section when the operational accuracy calibration
is complete.