Specifications

130
Chapter 6: Replacing Assemblies
To remove the logic analyzer cable
1 Remove power from the instrument
a Exit all logic analysis session. In the session manager, select Shutdown.
b At the query, select Power Down.
c When the “OK to power down” message appears, turn the instrument
off.
d Disconnect the power cord.
2 Remove the logic analyzer cable clamp.
a Remove two screws that secure the top logic analyzer cable clamp to
the outside rear panel.
b Remove two screws that secure the top cable clamp to the inside rear
panel.
c Slide the top cable clamp out of the rear panel.
3 Remove the logic analyzer cable.
a Gently lift the logic analyzer cable end connector from the circuit board
connector (J1 or J2).
b Remove the logic analyzer cable end connector through the rear panel.
The bottom cable clamp will also be removed with the logic analyzer
cable.
4 If the logic analyzer cable is faulty, replace the cable and follow the next
procedure to install the replacement logic analyzer cable.