Specifications
122
Chapter 5: Troubleshooting
2 Set up the stimulus board
a Configure the oscillator select switch S1 according to the following
settings:
•S1 Off
•S2 Off
•S3 Off
•Int
b Configure the data mode switch S4 according to the following settings:
•Even
•Count
c Press the Resynch VCO button, then the Counter RST (Counter Reset)
button.
3 Set up the logic analyzer
a Open the Session Manager window and select “Start Session.”
b In the Logic Analysis System window, select the module icon, then
select Setup and Trigger. A Setup and Trigger window appears.
c In the Logic Analysis System window, select the module icon, then
select Listing. A Listing window appears.
4 Set up the Sampling tab
a In the logic analyzer Setup and Trigger window, select the Sampling
tab.
b Under the Sampling tab, select State Mode.