Chapter 5: Troubleshooting
8 On the logic analyzer, select Run. The listing should look similar to the
figure below.
Scroll down at least 256 states to verify the data. The lower two bytes (four
digits) of Label1 show two incrementing binary counters. The upper two bytes of
Label1 show two decrementing binary counters. If the listing does not look
similar the figure, then there is a possible problem with the cable or high density
adapter. Causes for cable test failure include:
• open channel
• channel shorted to a neighboring channel
• channel shorted to either ground or a supply voltage
If the test data on either the lower two bytes or the upper two bytes is not
correct, then perform the following step to isolate the failure.
9 Verify the failure
a Reconnect probe adapter to the logic analyzer module. Connect logic
analyzer Pod 1 to the probe adapter output marked “Even.” Connect
logic analyzer Pod 2 to the probe adapter output marked “Odd.”