Specifications

113
Chapter 5: Troubleshooting
To test the cables using an Agilent E5378A Single-ended
Probe
This test allows you to functionally verify the logic analyzer cable and an Agilent
E5378A probe.
Equipment Required
1 Connect the logic analyzer to the stimulus board.
a Connect an Agilent E5378A 100-pin single-ended probe to the logic
analyzer module. Connect logic analyzer Pod 1 to the probe output
marked “Odd.” Connect logic analyzer Pod 2 to the probe output
marked “Even.”
b Connect the probe input to stimulus board connector Pod 1.
c Connect a power cord to the stimulus board power supply. Connect the
stimulus board power supply output to the stimulus board power
supply connector J82.
d Plug in the stimulus power supply into line power. The green LED DS1
Equipment Critical Specification Recommended Part
Stimulus Board No Substitute 16760-60001