Specifications

11
Chapter 1: General Information
Specifications
The specifications are the performance standards against which the product is
tested.
Setup/Hold Window:
tWidth
vHeight
Individual
Data Channel
tSample*
Clock Channel
vThreshold
0V
Sampling
Position
Data Eye
* User Adjustable
tSetup tHold
16760b05
Specifications for each input
Parameter
Minimum
Description/Notes
1500
Mb/s
1250
Mb/s
800
Mb/s
400
Mb/s
200
Mb/s
Data
to
Clock
tWidth* 500 ps 1.25 ns Eye width in system under test**
tSetup 250 ps 625 ps Data setup time required before tSample
tHold 250 ps 625 ps Data hold time required after tSample
*Specified for an input signal 800 mV peak-to-peak voltage swing 1V/ns slew rate, using an Agilent E5378A or E5382A probe.
**Eye width and height are specified at the probe tip. Eye width as measured by eye finder in the analyzer may be less.