Chapter 3: Testing Logic Analyzer Performance
Connect the Test Equipment
Verify and adjust 8133A pulse generator DC offset
1 On the 54845A oscilloscope, select Measure from the menu bar at the top
of the display.
2 Select Markers...
3 In the Markers Setup window set marker “Ay” to 0.7 V, and set marker
“By” to 1.3 V.
4 Observe the waveforms on the oscilloscope display. If they are not
centered within the “Ay” and “By” markers, adjust the 8133A pulse
generator’s Channel 1 OFFSET until the waveforms are centered as close
as possible. (The resolution of the 8133A OFFSET setting is 10 mV.)