Specifications
34
Chapter 3: Testing Logic Analyzer Performance
Connect the Test Equipment
5 Connect the E5383A Flying Lead Probe Set’s CLK lead to the pin strip of
the SMA/Flying Lead connector at the 8133A pulse generator’s Channel 1
OUTPUT.
NOTE: For each Flying Lead Probe connection, be sure to use a black ground lead
(five are supplied with the E5383A Flying Lead Probe Set) and orient the
leads so that the ground leads are connected to the SMA/Flying Lead
connector’s ground pins!
If you don’t have the ground leads, you can push the probe body’s ground
socket directly onto a ground pin on the SMA/Flying Lead test connector.
However, this will bend and could break pins on the SMA/Flying Lead test
connector because the probe body spacing is greater than the SMA/Flying
Lead test connector pin spacing. It is better to use the black ground leads.
In any case, the probe ground must be connected for each channel.
6 Connect the E5383A Flying Lead Probe Set’s bits 2 and 10 to the SMA/
Flying Lead test connector’s pin strip connector at the 8133A pulse
generator’s Channel 2 OUTPUT.
ground lead