Technical data

Troubleshooting
98 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
• "delay of channel [1|2] below minimum"
May happen:
If the frequency is decreased and the channel is in phase
mode, or
if the square mode is switched from square to another mode,
or
if clock mode is switched from internal/external direct to a
none direct mode.
• "delay of channel [1|2] above maximum"
May happen:
If the frequency is increased and the channel is in phase
mode, or
if the channel mode is switched from SQUARE to another
mode, or
if clock mode is switched from internal/external direct to a
none direct mode.
• "amplitude of trigger is too small"
May happen:
If the trigger high level is decreased, or
if the low level is increased.
• "amplitude of channel [1|2] is too small"
May happen:
If the channel high level is decreased, or
if the low level is increased.
• "amplitude of trigger is too large"
May happen:
If the trigger high level is increased, or
if the low level is decreased.
• "amplitude of channel [1|2] is too large"
May happen
If the channel high level is increased, or
if the low level is decreased.
• "low level of trigger is below minimum"
May happen: