Technical data
SCPI Commands Reference ARM Commands
94 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
:ARM[:SEQuence][:LAYer]:TERM
Syntax
:ARM[:SEQ][:LAY]:TERM[?]
Form Set & Query
Description Sets/reads the termination voltage for the start input signal.
Parameter Numeric [mV|V]
Specified limits –2.0 V … 3.0 V
*RST value 0 mV
Example Set the termination voltage to 1 V.
:ARM:TERM 1V
:ARM[:SEQuence][:LAYer][:STARt]
Syntax
:ARM[:SEQ][:LAY][STAR]
Form Event
Description Puts the instrument in armed mode. This means, that the
instrument waits for the selected edge to appear.
NOTE The instrument can only be put in armed mode if the external
start mode is selected (
“:ARM:SOURce EXT”)
.
Parameter –
*RST value –
Example Set the armed mode:
:ARM