Technical data
ARM Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 91
:ARM[:SEQuence][:LAYer]:LEVel
Syntax
:ARM[:SEQ][:LAY]:LEV
Form Set & Query
Description Specifies the threshold voltage for the external start signal.
NOTE The threshold can only be specified if the external start mode is
selected (
“:ARM:SOURce EXT”)
.
Parameter Numeric [mV|V]
Specified limits –2 V … 3.0 V
Absolute limits –2.0 V … 3.0 V
*RST value 100 mV
Example Sets the threshold voltage to 2.0 V.
:ARM:LEV 2.0 V