Technical data
SCPI Commands Reference ARM Commands
90 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
ARM Commands
The following table shows the Agilent 81133A/81134A Pulse
Generator ARM Commands:
The following figure shows how the ARM commands are
implemented in the 81133A/81134A user interface.
Command Parameter Description
:ARM
External start input
[:SEQuence]
[:LAYer]
:LEVel Numeric[mV|V]
Sets the trigger threshold
:SLOPe POSitive|NEGative
Trigger set to leading/trailing edge
of external signal
:SOURce IMMediate|
MANual|EXTernal
Sets the start input to
disabled/manual (by key)/external
started
:TERM[?] Numeric [mV|V]
Sets/reads termination voltage
[:STARt]
Starts signal output
:STOP
Stops signal output
:ARM[:SEQ][:LAY]:LEV
:ARM[:SEQ][:LAY]:TERM[?]
:ARM[:SEQ][:LAY]:SLOP
:ARM[:SEQ][:LAY][:START]
:ARM[:SEQ][:LAY]:SOURce
:ARM[:SEQ][:LAY]:STOP