Technical data
SYSTem Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 85
:SYSTem:COMMunicate:LAN[:SELF]
:DGATeway
Syntax
:SYST:COMM:LAN[:SELF]:DGAT[?]
Form Set & Query
Description Sets the instrument’s gateway.
NOTE This parameter must only be set if DHCP is not available. See
“:SYSTem:COMMunicate:LAN[:SELF]:DHCP” on page 83.
Parameter String <no>.<no>.<no>.<no> in quotes, where <no> is in the range 1
… 255.
*RST value –
Example
:SYST:COMM:LAN:DGAT "150.215.001.001"
:SYSTem:COMMunicate:GPIB[:SELF]:ADDR
Syntax
:SYST:COMM:GPIB[:SELF]:ADDR[?]
Form Set & Query
Description Sets/reads the instrument’s GPIB bus number.
Parameter Numeric
*RST value 13
Example
:SYST:COMM:GPIB[:SELF]:ADDR 15