Technical data

SCPI Commands Reference Status Handling Commands
76 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
:STATus:QUEStionable:FREQuency:NTRansition
Syntax
:STAT:QUES:FREQ:NTR[?]
Form Set & Query
Description Sets or queries the negative- transition register in the
QUEStionable:FREQency status group.
Parameter Numeric
*RST value
Specified limits 0 … 32767
:STATus:QUEStionable:FREQuency:PTRansition
Syntax
:STAT:QUES:FREQ:PTR[?]
FormForm Set & Query
Description Sets or queries the positive- transition register in the
QUEStionable:FREQency status group.
Parameter Numeric
*RST value
Specified limits 0 … 32767
:STATus:QUEStionable:MONotony
This command tree accesses the QUEStionable:MONotony status
group.
The QUEStionable:MONotony status group monitors the frequency,
width, delay and amplitude parameters. The range of these
parameters is made up of several internal ranges and when the
parameter moves from one internal range to the next a
discontinuity can occur. For example, increasing the frequency at a
range boundary could cause the actual output frequency to
decrease slightly. The range boundaries also vary with temperature,
and a significant temperature change could cause the instrument
to move to the next range in order to maintain the current
parameter within specification.