Technical data

SOURce Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 61
[:SOURce]:VOLTage[0|1|2][:LEVel]
[:IMMediate]:OFFSet
Syntax
[:SOUR]:VOLT[0|1|2][:LEV][:IMM]:OFFS[?]
Form Set & Query
Description Programs the offset of the output signal for the trigger output and
the channels.
Parameter Numeric [uV|mV|V] |MIN|MAX
Value coupling High = Offset + (Amplitude / 2)
Low = Offset – (Amplitude / 2)
Range coupling Amplitude
*RST value Trigger output (0): 0 V
Channels 1 and 2: 0 V
Specified limits Trigger channel (0): –1.975 V … +2.975 V
Channels 1 and 2: –1.975 V … +2.975 V
Example Set Trigger Output offset to –100 mV.
:VOLT0:OFFS -100MV