Technical data
OUTPut Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 43
:OUTPut[0|1|2][:STATe]
Syntax
:OUTP[0|1|2][:STAT][?]
Form Set & Query
Description Switches the trigger output and channel outputs on or off, where 0
is the trigger output.
For the two channel outputs, both OUTPUT and OUTPUT are
switched simultaneously. In query form, OFF is returned only if
both OUTPUT and
OUTPUT are off. They can be controlled
separately from the front panel, or by adding
:POS
or
:NEG
to the
command.
Parameter ON|OFF|1|0
*RST value OFF
Example Switch on the channel 1 outputs.
:OUTP1 ON
:OUTPut[1|2]:NEG
Syntax
:OUTP[1|2]:NEG[?]
Form Set & Query
Description Switches the specified channel OUTPUT on or off.
Parameter ON|OFF|1|0
*RST value OFF
Example Switch off the channel 1 OUTPUT.
:OUTP1:NEG OFF