Specifications
Table Of Contents
- General Information
- Preparing for Use
- To inspect package contents
- To connect power
- To connect the mouse, keyboard, LAN, printer, and GPIB cable
- To connect the standard 10073C probes
- To connect optional InfiniiMax oscilloscope probes
- To connect the digital probe
- Digital probe lead set
- To tilt the oscilloscope upward for easier viewing
- To turn on the oscilloscope
- To turn off the oscilloscope
- To verify basic oscilloscope operation
- Installing application programs on Infiniium
- Changing Windows System Settings
- To clean the oscilloscope
- Testing Performance
- Calibrating and Adjusting
- Troubleshooting
- To install the fan safety shield
- To troubleshoot the oscilloscope
- Primary Trouble Isolation
- No Display Trouble Isolation
- To check the backlight inverter voltages
- To check the display board video signals
- Power Supply Trouble Isolation
- To check probe power outputs
- To Check the keyboard
- To check the LEDs
- To check the motherboard, CPU, and RAM
- To setup the BIOS
- To troubleshoot the acquisition system
- Software Revisions
- Replacing Assemblies
- To return the oscilloscope to Agilent Technologies for service
- To remove and replace the top cover
- To remove and replace the bottom sleeve
- To disconnect and connect Mylar flex cables
- To remove and replace the CD-ROM drive
- To remove and replace the AutoProbe assembly
- To remove and replace the internal digital input cable (MSO models only)
- To remove and replace the backlight inverter board
- To remove and replace the front panel assembly
- To remove and replace the keyboard, touch screen, and flat-panel display assemblies
- To remove and replace the acquisition board assembly
- To remove and replace the PCI bridge board
- To remove and replace the display board
- To remove and replace the hard disk drive
- To remove and replace the motherboard
- To replace the Intel motherboard with the ADLINK motherboard
- To remove and replace the power supply
- To remove and replace the fan controller board
- To remove and replace a fan
- To remove and replace the probe power and control assembly
- Replaceable Parts
- Theory of Operation
- Index

Chapter 5: Troubleshooting
To troubleshoot the acquisition system
93
Figure 5-12
Acquisition Trouble Isolation
Acquisition Trouble Isolation
Does
vertical test
group pass?
Does
trigger test group
pass?
Yes
Does
TimeBase test
group pass?
Yes
Does
ADC test group
pass?
Yes
Does
acquisition memory
test group
pass?
Yes
Does
acquisition interface
test group
pass?
Does
misc. scope
test group
pass?
Yes
Yes
Replace
acquisition
board.
Replace the interface board, A10. If
that does not work, replace the
display board.
Does
temp sense
pass?
Does
fan status
pass?
Check fans
Replace fans
that are not
turning, if
none are
turning
replace
power board
Does
NVram tests
pass?
Replace
interface
card.
Go to 'Primary
Trouble Isolation'.
No
No
Yes
Yes
Replace
acquisition
board.
No
No
No
No
No
No
No
No
Go to 'Primary
Trouble Isolation'.
Yes
Yes