Specifications
Table Of Contents
- General Information
- Preparing for Use
- To inspect package contents
- To connect power
- To connect the mouse, keyboard, LAN, printer, and GPIB cable
- To connect the standard 10073C probes
- To connect optional InfiniiMax oscilloscope probes
- To connect the digital probe
- Digital probe lead set
- To tilt the oscilloscope upward for easier viewing
- To turn on the oscilloscope
- To turn off the oscilloscope
- To verify basic oscilloscope operation
- Installing application programs on Infiniium
- Changing Windows System Settings
- To clean the oscilloscope
- Testing Performance
- Calibrating and Adjusting
- Troubleshooting
- To install the fan safety shield
- To troubleshoot the oscilloscope
- Primary Trouble Isolation
- No Display Trouble Isolation
- To check the backlight inverter voltages
- To check the display board video signals
- Power Supply Trouble Isolation
- To check probe power outputs
- To Check the keyboard
- To check the LEDs
- To check the motherboard, CPU, and RAM
- To setup the BIOS
- To troubleshoot the acquisition system
- Software Revisions
- Replacing Assemblies
- To return the oscilloscope to Agilent Technologies for service
- To remove and replace the top cover
- To remove and replace the bottom sleeve
- To disconnect and connect Mylar flex cables
- To remove and replace the CD-ROM drive
- To remove and replace the AutoProbe assembly
- To remove and replace the internal digital input cable (MSO models only)
- To remove and replace the backlight inverter board
- To remove and replace the front panel assembly
- To remove and replace the keyboard, touch screen, and flat-panel display assemblies
- To remove and replace the acquisition board assembly
- To remove and replace the PCI bridge board
- To remove and replace the display board
- To remove and replace the hard disk drive
- To remove and replace the motherboard
- To replace the Intel motherboard with the ADLINK motherboard
- To remove and replace the power supply
- To remove and replace the fan controller board
- To remove and replace a fan
- To remove and replace the probe power and control assembly
- Replaceable Parts
- Theory of Operation
- Index

Chapter 5: Troubleshooting
To troubleshoot the oscilloscope
75
Primary Trouble Isolation Flowchart
Primary Trouble Isolation
Perform power-up.
Check display.
On
screen display
problems
?
Run scope self tests.
Does
self test
pass?
Check front panel
response.
Do
knobs and keys test
OK?
System works; do
performance tests.
End
Go to 'No Display
Debug.'
Go to 'Keyboard
Troubleshooting.'
Does
LED test
OK?
Check Self Calibration.
Does
self calibration test
pass?
Go to 'Acquisition
Troubleshooting.'
No
No
No
Yes
A
B
E
F
G
H
54830F01a
Check processor
temperature.
C
Is
temperature
OK?
Yes
Check for fan fail
message.
Does
fan fail?
No
Replace fan.
Yes
D
Go to 'Acquisition
Troubleshooting.'
Yes
No
Yes
No
Yes
Yes
Replace
motherboard.
Replace keypad
(includes LEDs).
No