Specifications
Table Of Contents
- General Information
- Preparing for Use
- To inspect package contents
- To connect power
- To connect the mouse, keyboard, LAN, printer, and GPIB cable
- To connect the standard 10073C probes
- To connect optional InfiniiMax oscilloscope probes
- To connect the digital probe
- Digital probe lead set
- To tilt the oscilloscope upward for easier viewing
- To turn on the oscilloscope
- To turn off the oscilloscope
- To verify basic oscilloscope operation
- Installing application programs on Infiniium
- Changing Windows System Settings
- To clean the oscilloscope
- Testing Performance
- Calibrating and Adjusting
- Troubleshooting
- To install the fan safety shield
- To troubleshoot the oscilloscope
- Primary Trouble Isolation
- No Display Trouble Isolation
- To check the backlight inverter voltages
- To check the display board video signals
- Power Supply Trouble Isolation
- To check probe power outputs
- To Check the keyboard
- To check the LEDs
- To check the motherboard, CPU, and RAM
- To setup the BIOS
- To troubleshoot the acquisition system
- Software Revisions
- Replacing Assemblies
- To return the oscilloscope to Agilent Technologies for service
- To remove and replace the top cover
- To remove and replace the bottom sleeve
- To disconnect and connect Mylar flex cables
- To remove and replace the CD-ROM drive
- To remove and replace the AutoProbe assembly
- To remove and replace the internal digital input cable (MSO models only)
- To remove and replace the backlight inverter board
- To remove and replace the front panel assembly
- To remove and replace the keyboard, touch screen, and flat-panel display assemblies
- To remove and replace the acquisition board assembly
- To remove and replace the PCI bridge board
- To remove and replace the display board
- To remove and replace the hard disk drive
- To remove and replace the motherboard
- To replace the Intel motherboard with the ADLINK motherboard
- To remove and replace the power supply
- To remove and replace the fan controller board
- To remove and replace a fan
- To remove and replace the probe power and control assembly
- Replaceable Parts
- Theory of Operation
- Index

Chapter 3: Testing Performance
To test delta time measurement accuracy
63
Procedure
1
Configure the sine wave source to output a 0 dBm (600 mVpp) sine wave into 50 ohms
with a frequency that corresponds to the table below.
2 Press the Default Setup key on the oscilloscope.
3 Set the scope's input impedance to 50 ohms.
4 Set the channel 1 vertical scale to 100 mV/div.
5 Set the scope's horizontal scale to 5 ns/div.
6 If necessary, adjust the trigger level until the oscilloscope triggers without auto-
triggering.
7 Enable waveform averaging, with 16 averages.
8 Set the measurement thresholds for all waveforms to a fixed voltage level of 0 V and
+- 20 mV hysteresis.
9 Enable a peak-to-peak voltage measurement on channel 1.
10 Connect the sine wave source through the appropriate band-pass filter to the
oscilloscope's channel 1 input, as shown in the diagram above. The appropriate band-
pass filter can be found from the table below for each scope model.
11 Adjust the sine wave amplitude until the peak-to-peak voltage of the displayed channel
1 waveform equals 500 mV +- 5 mV. Press
Clear Display after each adjustment to clear
accumulated averages.
12 Turn off the peak-to-peak voltage measurement.
13 Set the number of waveform averages to 256.
14 Set the acquisition memory depth to 65536 points.
15 On the oscilloscope, press Stop.
16 Press Clear Display.
17 Press Run, wait until 256 averages have accumulated, and then press Stop.
18 Enable jitter mode.
19 Enable period measurements on channel 1 using rising edges.
20 Record the minimum, maximum, and standard deviation values for the period
measurement on channel 1. Compare these values to the test limits for averaged mode
DTMA shown in the Performance Test Record.
21 Disable waveform averaging.
22 Set the trigger mode to single.
23 Press Clear Display.
24 Press Run.
25 Record the minimum, maximum, and standard deviation values for the period
measurement on channel 1. Compare these values to the test limits for non-averaged
mode DTMA shown in the Performance Test Record.
26 Repeat steps 2 through 25 for channels 2, 3, and 4.
Oscilloscope Model Sine Wave Frequency
8104 501 MHz
8064 301 MHz
Oscilloscope Model Band-pass Filter Center Frequency
8104 500 MHz
8064 300 MHz