Specifications
Table Of Contents
- General Information
- Preparing for Use
- To inspect package contents
- To connect power
- To connect the mouse, keyboard, LAN, printer, and GPIB cable
- To connect the standard 10073C probes
- To connect optional InfiniiMax oscilloscope probes
- To connect the digital probe
- Digital probe lead set
- To tilt the oscilloscope upward for easier viewing
- To turn on the oscilloscope
- To turn off the oscilloscope
- To verify basic oscilloscope operation
- Installing application programs on Infiniium
- Changing Windows System Settings
- To clean the oscilloscope
- Testing Performance
- Calibrating and Adjusting
- Troubleshooting
- To install the fan safety shield
- To troubleshoot the oscilloscope
- Primary Trouble Isolation
- No Display Trouble Isolation
- To check the backlight inverter voltages
- To check the display board video signals
- Power Supply Trouble Isolation
- To check probe power outputs
- To Check the keyboard
- To check the LEDs
- To check the motherboard, CPU, and RAM
- To setup the BIOS
- To troubleshoot the acquisition system
- Software Revisions
- Replacing Assemblies
- To return the oscilloscope to Agilent Technologies for service
- To remove and replace the top cover
- To remove and replace the bottom sleeve
- To disconnect and connect Mylar flex cables
- To remove and replace the CD-ROM drive
- To remove and replace the AutoProbe assembly
- To remove and replace the internal digital input cable (MSO models only)
- To remove and replace the backlight inverter board
- To remove and replace the front panel assembly
- To remove and replace the keyboard, touch screen, and flat-panel display assemblies
- To remove and replace the acquisition board assembly
- To remove and replace the PCI bridge board
- To remove and replace the display board
- To remove and replace the hard disk drive
- To remove and replace the motherboard
- To replace the Intel motherboard with the ADLINK motherboard
- To remove and replace the power supply
- To remove and replace the fan controller board
- To remove and replace a fan
- To remove and replace the probe power and control assembly
- Replaceable Parts
- Theory of Operation
- Index

Chapter 3: Testing Performance
To verify threshold accuracy
55
High-to-Low Transitions
1 Set the DC Signal Source voltage to the "Initial V1 Setting" as shown in the table below.
The DC Signal Source voltage is referred to as "V1" throughout this procedure.
Table 1: Oscilloscope Threshold Voltage Setting and Initial V1 Setting
Note: In this test, complete steps 2, 3 and 4 for the first threshold voltage setting in the table
above. Then set the threshold voltage to the next value in the table and repeat steps 2, 3 and 4
for this threshold voltage setting. Finally, set the threshold voltage to the last value in the table
and repeat steps 2, 3 and 4 for the last threshold voltage setting.
2
Check the logic display at the bottom of the screen to ensure that all bits are logic 1.
If this is not the case, check the following:
• Read the DMM to determine if the input voltage is correctly set to the "Initial V1 Setting" value
shown in Table 1. Adjust the DC Signal Source voltage if necessary.
• Check the oscilloscope’s Digital set up to ensure that the logic threshold voltage is set to the
"Oscilloscope Threshold Voltage Setting" value shown in Table 1, and ensure that the display
of all bits is enabled.
• Ensure that the oscilloscope is in Run mode.
• Ensure that the oscilloscope trigger system is properly configured.
• If the oscilloscope is set up correctly and the input voltage is accurately set to the initial V1
setting and at least one bit is displayed as 0, then either the oscilloscope is out of calibration
or it is malfunctioning. If the oscilloscope has been powered on for at least 30 minutes, re-
calibrate the oscilloscope and retry this set up.
3
Perform the coarse threshold voltage search (to determine the measured threshold
voltage within ±50mV.) You will repeatedly adjust the voltage until all bits change from
logic 1 to logic 0. Record the voltage you applied just before all bits switched.
a Repeat the following steps until all displayed bits are logic 0:
• Reduce V1 by 50mV and examine the logic display.
• If all bits are 0, record the initial value of V1 (the value applied just before all bits
switched) as the V2 value in Table 2 and proceed to the Fine Threshold Voltage search.
• If at least one bit is still 1, reset the input voltage to the initial value of V1 shown in
Table 1. This is done so that when V1 is reduced, a significant voltage transition occurs.
• Reduce V1 by 100mV and examine the logic display.
• If all bits are 0, record the previous value of V1 (the value at which at least one logic bit
was still logic 1) as V2 in the table below. Proceed to the Fine Threshold Voltage search.
Oscilloscope Threshold Voltage Setting Initial V1 Setting
+5V 5.35V
0V 200mV
-5V -4.65V