Specifications
Table Of Contents
- General Information
- Preparing for Use
- To inspect package contents
- To connect power
- To connect the mouse, keyboard, LAN, printer, and GPIB cable
- To connect the standard 10073C probes
- To connect optional InfiniiMax oscilloscope probes
- To connect the digital probe
- Digital probe lead set
- To tilt the oscilloscope upward for easier viewing
- To turn on the oscilloscope
- To turn off the oscilloscope
- To verify basic oscilloscope operation
- Installing application programs on Infiniium
- Changing Windows System Settings
- To clean the oscilloscope
- Testing Performance
- Calibrating and Adjusting
- Troubleshooting
- To install the fan safety shield
- To troubleshoot the oscilloscope
- Primary Trouble Isolation
- No Display Trouble Isolation
- To check the backlight inverter voltages
- To check the display board video signals
- Power Supply Trouble Isolation
- To check probe power outputs
- To Check the keyboard
- To check the LEDs
- To check the motherboard, CPU, and RAM
- To setup the BIOS
- To troubleshoot the acquisition system
- Software Revisions
- Replacing Assemblies
- To return the oscilloscope to Agilent Technologies for service
- To remove and replace the top cover
- To remove and replace the bottom sleeve
- To disconnect and connect Mylar flex cables
- To remove and replace the CD-ROM drive
- To remove and replace the AutoProbe assembly
- To remove and replace the internal digital input cable (MSO models only)
- To remove and replace the backlight inverter board
- To remove and replace the front panel assembly
- To remove and replace the keyboard, touch screen, and flat-panel display assemblies
- To remove and replace the acquisition board assembly
- To remove and replace the PCI bridge board
- To remove and replace the display board
- To remove and replace the hard disk drive
- To remove and replace the motherboard
- To replace the Intel motherboard with the ADLINK motherboard
- To remove and replace the power supply
- To remove and replace the fan controller board
- To remove and replace a fan
- To remove and replace the probe power and control assembly
- Replaceable Parts
- Theory of Operation
- Index

Chapter 3: Testing Performance
To test the DC calibrator
41
5
Set the DC output voltage to +2.400 V using the Level spin box or the numeric keypad
dialog.
You can access the numeric keypad dialog by selecting (clicking on) the value in the Level box.
Enter the values by selecting digits, signs, and exponents in the keypad. Select Close when
finished.
6
The DVM should read near +2.400 V. Record the reading to four significant digits.
V1 = __________.
7 Set the level to -2.400 V using the Level spin box or the numeric keypad dialog.
8 The DVM should read near –2.400 V. Record the reading. V2 = __________.
9 Subtract the second reading from the first reading, then divide the result by 4.8 V.
For example, if V1 is +2.397 V and V2 is -2.395 V, then
10
The final result should be between 0.998 and 1.002. Record the result in the
"Performance Test Record" on page 64.
11 Select Close to exit the calibration menu.
If the test fails
Repair is necessary. See Chapter 5, “Troubleshooting.”
V1-V2
4.8 V
----------------
2.397 V-(-2.395 V)
4.8 V
---------------------------------------------- 0 . 9 9 8==