Specifications

24
Infiniium Performance Characteristics (continued)
Acquisition: Scope channels DSO8064A, MSO8064A, DSO8104A, MSO8104A
Real time sample rate (max)
2 channels 4 GSa/s
Each channel 2 GSa/s
Equivalent time sample rate (max)
250 GSa/s
Memory depth
6
2 channels/4 channels
Standard 8 Mpts/4 Mpts
Option 080 16 Mpts/8 Mpts
Option 160 32 Mpts/16 Mpts
Option 320 64 Mpts/32 Mpts
Option 640 128 Mpts/64 Mpts
Sampling modes
Real time
Normal Successive single-shot acquisitions
Peak detect Captures and displays narrow pulses or glitches at all real time sample rates
High resolution Real-time boxcar averaging reduces random noise and increases resolution
Equivalent time Random repetitive sampling (higher time resolution at faster sweep speeds)
Segmented memory Captures bursting signals at maximum sample rate without consuming memory during
periods of inactivity. Selectable number of segments up to 32,768 depending on
memory option installed. Minimum inter-segment time (or the time between the end of
the previous acquisition and the beginning of the next acquisition) of 20 μs.
Averaging Selectable from 2 to 65,534
Filters (Sin[x])/x Interpolation Filter on/off selectable FIR digital filter. Digital signal processing adds points between
acquired data points to enhance measurement accuracy and waveform display quality.
BW = sample rate/4
Acquisition: Digital channels MSO8064A, MSO8104A
Maximum input frequency 250 MHz
Maximum real time sample rate 1 GSa/s
Memory depth per channel 32 M
Minimum width glitch detection 2.5 ns
6. Maximum 2-channel memory depth only available at maximum 2-channel sample rate. Maximum each channel memory depth
available at any selectable sample rate.