Specifications
23
Infiniium Performance Characteristics (continued)
Horizontal DSO8064A, MSO8064A, DSO8104A, MSO8104A
DSO8064A/MSO8064A DSO8104A/MSO8104A
Main time base range 500 ps/div to 20 s/div 200 ps/div to 20 s/div
Horizontal position range 0 to ± 200 s
Delayed sweep range 1 ps/div to current main time base setting
Resolution 4 ps
Time scale accuracy
1
±(15+2•(yearssincemanufacture))ppm
Delta-time measurement accuracy
1
Values apply to period measurements of 0.5 Vpp sine waves with frequencies equal to
½ the oscilloscope bandwidth at 100 mV/div using ≥ 2 GSa/s and sinx/x interpolation
≥ 256 averages, absolute (ps peak) 4.0 2.0
≥ 256 averages, standard deviation (ps rms) 1.2 0.38
No averaging, absolute (ps peak) 30 27
No averaging, standard deviation 6.8 5.7
Channel-to-channel deskew range –100 μs to 100 μs
Modes Main, delayed, roll
Reference positions Left, center, right
Jitter measurement floor
Time interval error 7 ps rms 5 ps rms
Period jitter 10 ps rms 7 ps rms
N-cycle, cycle-cycle jitter 15 ps rms 11 ps rms
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30 minute warm-up period and ± 10 °C from
firmware calibration temperature.