consists of sending the *IDN? command and checking the response. This
test can be used to verify that the module is connected properly and is
responding to a basic command.
Procedure 1. Verify that the module is properly installed in mainframe
2. Verify that the mainframe has passed its power-on test.
3. Send *IDN? to the module (see example following)
4. The return should be as follows (revision number may vary):
HEWLETT-PACKARD,E1330A,0,A.06.00
NOTES
If the primary address setting, secondary address setting, or the interface
select code is set incorrectly, the module will not respond. Verify proper
address selection before troubleshooting.
Both the Agilent E1330A and Agilent E1330B return the string shown in
Step 4. The E1330B will return "E1330A" in response to the *IDN? query.
Example An example follows which uses an HP 9000 Series 300 computer with
BASIC and a module address of 70918.
10 DIM A$[100]
20 OUTPUT 70918;"*IDN?" Send the ID command
30 ENTER 70918;A$ Get response
40 PRINT A$
50 END
Operation
Verification
Test
The procedures in this section are used to provide a high level of confidence
that the module is meeting published specifications. The Operation
Verification test is a subset of the Performance Verification tests and is
suitable for checkout after performing repairs.
The Operation Verification Test is performed by completing the Digital Test
(Test 2-1) as described in the Performance Verification test procedures. This
test is usually sufficient to verify that the module is meeting its
specifications.