Mainframes User's Manual
Agilent E1333A 3-Channel Universal
Counter Menu
Menu Levels and Content
Level 1 Level 2 Level 3 Level 4 Level 5 User Entry  Command( s)  Used  Description 
COUNTER MONITOR CHAN1 DISP:MON:CHAN 1;STAT ON  Monitor instrument operations on channel 1 
CHAN2 DISP:MON:CHAN 2;STAT ON  Monitor instrument operations on channel 2 
CHAN3 DISP:MON:CHAN 3;STAT ON  Monitor instrument operation on channel 3 
AUTO DISP:MON:CHAN AUTO;STAT ON  Monitor instrument operations on active channel 
INPUT LEVEL CHAN1 voltage †  SENS1:EVEN:LEV< value>  Set trigger level voltage for channel 1
CHAN2 voltage †  SENS2:EVEN:LEV< value>  Set trigger level voltage for channel 2
SLOPE CH AN1 POS SENS1: EVEN: SLOP POS  Posi t i ve t r i gger  sl ope f or  chan n el  1
NEG SENS1:EVEN:SLOP NEG  Negative trigger slope for channel 1
CHAN2 POS SENS2: EVEN:SLOP POS  Posit ive t r i gger slope for  channel 2
NEG SENS2:EVEN:SLOP NEG  Negative trigger slope for channel 2
COUPLE AC INP:COUP AC  AC-coupled input (channels 1 & 2 only)
DC INP:COUP DC  DC-coupled input (channels 1&2)
IMPED 50_OHM INP:IMP 50  50Ω input resistance (channels 1 & 2 only)
1_MOHM INP:IMP 1e6  1MΩ input resistance (channels 1 & 2 only)
ATTEN 0dB INP:ATT 0  No input attenuation (channels 1 & 2 only)
20dB INP:ATT 20  20dB input attenuation (channels 1 & 2 only)
FILTER ON INP:FILT ON  Input filter on (channels 1 & 2 only)
OFF INP:FILT OFF  Input filter off (channels 1 & 2 only)
FREQ CHAN1 TRIG:SOUR IMM;:MEAS1:FREQ?  Frequency measurement on channel 1
CHAN2 TRIG:SOUR IMM;:MEAS2:FREQ?  Frequency measurement on channel 2 
CHAN3 TRIG:SOUR IMM;:MEAS3:FREQ?  Frequency measurement on channel 3 
PERIOD CHAN1 TRIG:SOUR IMM;:MEAS1:PER?  Period measurement on channel 1
CHAN2 TRIG:SOUR IMM;:MEAS2:PER?  Period measurement on channel 2 
(continued on following page)
2-26 Using the Display Terminal Interface